Importance of exercise testing shortly after subcutaneous implantable cardioverter-defibrillator implantation in patients with Brugada syndrome – The first case of associated inappropriate shock in Japan

Abstract We report the case of a 51-year-old patient with Brugada syndrome (BrS) who experienced inappropriate shock due to T-wave oversensing (TWOS) during exercise when the optimal sensing vector was selected based on the automatic analysis by a subcutaneous implantable cardioverter-defibrillator...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Journal of arrhythmia 2017-04, Vol.33 (2), p.156-158
Hauptverfasser: Ishibashi, Kohei, MD, PhD, Noda, Takashi, MD, PhD, Kamakura, Tsukasa, MD, PhD, Wada, Mitsuru, MD, Inoue, Yuko, MD, PhD, Okamura, Hideo, MD, PhD, Kusano, Kengo, MD, PhD
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext bestellen
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:Abstract We report the case of a 51-year-old patient with Brugada syndrome (BrS) who experienced inappropriate shock due to T-wave oversensing (TWOS) during exercise when the optimal sensing vector was selected based on the automatic analysis by a subcutaneous implantable cardioverter-defibrillator (S-ICD). After selecting another vector during exercise testing, TWOS did not re-occur. Selection of appropriate sensing vector based on analyses under various conditions, including during exercise after S-ICD implantation, should be considered for patients with BrS.