Pragmatic Micrometre to Millimetre Calibration Using Multiple Methods for Low-Coherence Interferometer in Embedded Metrology Applications
In-situ metrology utilised for surface topography, texture and form analysis along with quality control processes requires a high-level of reliability. Hence, a traceable method for calibrating the measurement system’s transfer function is required at regular intervals. This paper compares three met...
Gespeichert in:
Veröffentlicht in: | Sensors (Basel, Switzerland) Switzerland), 2021-07, Vol.21 (15), p.5101 |
---|---|
Hauptverfasser: | , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | In-situ metrology utilised for surface topography, texture and form analysis along with quality control processes requires a high-level of reliability. Hence, a traceable method for calibrating the measurement system’s transfer function is required at regular intervals. This paper compares three methods of dimensional calibration for a spectral domain low coherence interferometer using a reference laser interferometer versus two types of single material measure. Additionally, the impact of dataset sparsity is shown along with the effect of using a singular calibration dataset for system performance when operating across different media. |
---|---|
ISSN: | 1424-8220 1424-8220 |
DOI: | 10.3390/s21155101 |