Bioreactor design for the mechanical stimulation by compression of 3D cell cultures
Bioreactors with a controlled physiological environment are being developed to study various cell processes. The influences of mechanostimulation on bone cell cultures can be investigated using a compression bioreactor. The developed bioreactor system applies a cyclic compression force to the specim...
Gespeichert in:
Veröffentlicht in: | Current directions in biomedical engineering 2021-10, Vol.7 (2), p.899-902 |
---|---|
Hauptverfasser: | , , , , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Bioreactors with a controlled physiological environment are being developed to study various cell processes. The influences of mechanostimulation on bone cell cultures can be investigated using a compression bioreactor. The developed bioreactor system applies a cyclic compression force to the specimen via an eccentrically mounted push rod. The compression force is monitored by a force sensor to detect changes in the material properties of the specimen. Depending on the piston setting, a stroke of 0.28 - 2.50 mm can be applied to the specimen. The bioreactor system was tested with a trial run of 18 days. A sample was continuously stimulated with a loading frequency of 2 Hz and a stroke of 1.50 mm. The sterility in the cell chamber as well as the functionality of the realised bioreactor stimulation system could be successfully confirmed |
---|---|
ISSN: | 2364-5504 2364-5504 |
DOI: | 10.1515/cdbme-2021-2229 |