Metallic allergy requiring removal of iliac stent: report of a case

Background The use of metallic stents, which are predominantly composed of nickel, in the treatment of patients with nickel allergy has not been well studied. Case presentation A patient who suffered from contact dermatitis due to implantation of an iliac stent was successfully treated by removing t...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Surgical Case Reports 2020-04, Vol.6 (1), p.82-82, Article 82
Hauptverfasser: Guntani, Atsushi, Kawakubo, Eisuke, Yoshiga, Ryosuke, Mii, Shinsuke
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:Background The use of metallic stents, which are predominantly composed of nickel, in the treatment of patients with nickel allergy has not been well studied. Case presentation A patient who suffered from contact dermatitis due to implantation of an iliac stent was successfully treated by removing the stent that caused nickel allergy. The patient has exhibited no symptoms of claudication or severe pruritic rash in the 2-year follow-up period after iliac stent removal. Conclusions We herein report a case of nickel allergy in which a metallic iliac stent that was predominantly composed of nickel was removed. The patient showed a marked recovery from her contact dermatitis. In cases of suspected metallic allergy, it is necessary to consider revascularization without using a metal device.
ISSN:2198-7793
2198-7793
DOI:10.1186/s40792-020-00843-8