Influence of jaw setting in the determination of stereotactic small-field output factors with different detectors

Background: The experimental determination of relative output factors presents the greatest challenge, especially for small fields with different detectors. The aim of this study is to evaluate the influence of jaw positions on small-field output factors for the fields defined by micro-multileaf col...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Journal of medical physics 2022-01, Vol.47 (1), p.65-72
Hauptverfasser: George, Seby, Ponmalar, Y, Godson, Henry, Kumar, A, Ravindran, B
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:Background: The experimental determination of relative output factors presents the greatest challenge, especially for small fields with different detectors. The aim of this study is to evaluate the influence of jaw positions on small-field output factors for the fields defined by micro-multileaf collimator and circular cones with different detectors. Materials and Methods: The stereotactic output factors were measured on Primus linear accelerator with BrainLab micro-multileaf collimator (mMLC) and circular cones as add-on tertiary collimators. Square field sizes ranging from 0.6 cm × 0.6 cm to 9.8 cm × 9.8 cm and circular fields of diameter ranging from 1.0 cm to 4.0 cm were defined by mMLC and circular cones, respectively. The influence of jaw position on output factor was assessed for different geometric configurations with three different detectors. Results: The values obtained with PinPoint ion chamber were consistent with microDiamond detector for fields greater than 24 mm × 24 mm, but an underestimation of 23.9% was noticed in 6 mm x 6 mm field size. For the mMLC defined field size of 6 mm × 6 mm, when the X-Y jaw was moved from 8 mm × 8 mm to 80 mm × 80 mm, an increase in the output by a factor of 1.7 was observed with both microDiamond and stereotactic radiosurgery diode, whereas an increase in output by a factor of 1.9 was noticed with PinPoint ion chamber. Conclusion: Output factors obtained with different detectors show high differences in the smallest field size for all collimating systems. This study confirms that the position of X and Y jaw above the tertiary collimator significantly influences the small-field output factor.
ISSN:0971-6203
1998-3913
DOI:10.4103/jmp.jmp_111_21