A prospective study of contrast preservation using ultra-low contrast delivery technique versus standard automated contrast injector system in coronary procedures
We aimed to assess the decrease in contrast media volume (CMV) with ultra-low contrast delivery technique (ULCD) developed at our institution versus the usual automated contrast injector system (ACIS) contrast delivery in coronary procedures. We analyzed the amount of contrast given in the consecuti...
Gespeichert in:
Veröffentlicht in: | Indian heart journal 2019-07, Vol.71 (4), p.297-302 |
---|---|
Hauptverfasser: | , , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | We aimed to assess the decrease in contrast media volume (CMV) with ultra-low contrast delivery technique (ULCD) developed at our institution versus the usual automated contrast injector system (ACIS) contrast delivery in coronary procedures.
We analyzed the amount of contrast given in the consecutive 204 patients of the operators who use ULCD technique versus consecutive 200 patients of the other operators who use ACIS without ULCD technique for coronary angiograms and/or percutaneous coronary interventions (PCIs) from May 2017 to July 2018 at our center. We calculated the mean CMV between these groups.
We observed a significant reduction in mean CMV with ULCD technique versus standard ACIS, respectively: angiogram 24.8 ± 15.8 mL (n = 194) vs 42.3 ± 25.1 mL (n = 200) (p |
---|---|
ISSN: | 0019-4832 2213-3763 |
DOI: | 10.1016/j.ihj.2019.09.001 |