MONITORING OF INDIVIDUAL DOSES FOR MEDICAL WORKERS OF DENTAL POLYCLINIC’S X-RAY ROOMS IN DUSHANBE, THE REPUBLIC OF TADJIKISTAN

The article presents the data and analyses of personnel’s average annual external exposure doses monitoring via the thermoluminescent dosimetry method used for X-ray radiological personnel in dental polyclinics of Dushanbe, Tadjikistan Republic over a 5-year period ( 2010–2014 ). Out of 42 registere...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Радиационная гигиена 2016-04, Vol.9 (1), p.58-60
Hauptverfasser: N. U. Hakimova, E. Yu. Malysheva, Sh. G. Shosafarova, U. M. Mirsaidov
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:The article presents the data and analyses of personnel’s average annual external exposure doses monitoring via the thermoluminescent dosimetry method used for X-ray radiological personnel in dental polyclinics of Dushanbe, Tadjikistan Republic over a 5-year period ( 2010–2014 ). Out of 42 registered medical institutions dental polyclinics amounted up to only just 14%. For this work thermoluminescent dosimeters were used ( with LiF: Mg, Ti ) with the thermoluminescent dosimetric installation “ Harshaw – 4500” as the reader device. Monitoring results comparison of individual dose equivalent Hp ( 10 ) values was conducted for two groups of medical workers: medical doctors and X-ray lab technicians. It is demonstrated that radiological technicians’ professional exposure doses are on the average by 23% higher than those for medical doctors.The average individual exposure doses over the above indicated period amount to 0,93 mSv and 1,3 mSv for doctors and X-ray lab technicians, respectively, and are in the range from 0,45 mSv to 2,39 mSv. The doses include contribution from the natural background. The values of doses recorded for the personnel in dental polyclinic correspond to those recorded for the workers in the routine X-ray rooms.
ISSN:1998-426X