Inappropriate defibrillator shock due to EMI despite noise reversion and SecureSense - What is the mechanism?

Implantable cardioverter defibrillator (ICD) shocks are always a matter of anxiety and concern for both patient and clinician. We report a case of ischemic cardiomyopathy who reported to us after receiving two shocks within 48 hours. The therapy was inappropriate as noted from device logs. EMI was i...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Indian pacing and electrophysiology journal 2022-03, Vol.22 (2), p.112-114
Hauptverfasser: Mukherjee, Sanjeev S, Bera, Debabrata, Kar, Ayan
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:Implantable cardioverter defibrillator (ICD) shocks are always a matter of anxiety and concern for both patient and clinician. We report a case of ischemic cardiomyopathy who reported to us after receiving two shocks within 48 hours. The therapy was inappropriate as noted from device logs. EMI was identified as the incriminating cause. We discuss the reasons why he received the therapy despite active ‘Noise reversion’ and ‘SecureSense’ algorithms to avoid noise /EMI related oversensing.
ISSN:0972-6292
0972-6292
DOI:10.1016/j.ipej.2022.01.006