Comprehensive hemocompatibility analysis on the application of diamond-like carbon to ePTFE artificial vascular prosthesis

The aim of this study was to obtain comprehensive data regarding the hemocompatibility of diamond-like carbon (DLC)-coated expanded polytetrafluoroethylene (ePTFE). DLC increased the hydrophilicity and smoothened the surface and fibrillar structure, respectively, of the ePTFE. DLC-coated ePTFE had m...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Scientific reports 2023-05, Vol.13 (1), p.8386-8386, Article 8386
Hauptverfasser: Goyama, Takashi, Fujii, Yasuhiro, Muraoka, Genya, Nakatani, Tatsuyuki, Ousaka, Daiki, Imai, Yuichi, Kuwada, Noriaki, Tsuji, Tatsunori, Shuku, Takayuki, Uchida, Haruhito A., Nishibori, Masahiro, Oozawa, Susumu, Kasahara, Shingo
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:The aim of this study was to obtain comprehensive data regarding the hemocompatibility of diamond-like carbon (DLC)-coated expanded polytetrafluoroethylene (ePTFE). DLC increased the hydrophilicity and smoothened the surface and fibrillar structure, respectively, of the ePTFE. DLC-coated ePTFE had more albumin and fibrinogen adsorption and less platelet adhesion than uncoated ePTFE. There were scarce red cell attachments in in vitro human and in vivo animal (rat and swine) whole blood contact tests in both DLC-coated and uncoated ePTFE. DLC-coated ePTFE had a similar but marginally thicker band movement than uncoated-ePTFE with SDS-PAGE after human whole blood contact test. In addition, survival studies of aortic graft replacement in rats (1.5 mm graft) and arteriovenous shunt in goats (4 mm graft) were performed to compare the patency and clot formation between DLC-coated and uncoated ePTFE grafts. Comparable patency was observed in both animal models. However, clots were observed in the luminal surface of the patent 1.5 mm DLC-coated ePTFE grafts, but not in that of uncoated ePTFE grafts. In conclusions, hemocompatibility of DLC-coated ePTFE was high and comparable to that of uncoated ePTFE. However, it failed to improve the hemocompatibility of 1.5 mm ePTFE graft probably because increased fibrinogen adsorption canceled the other beneficial effects of DLC.
ISSN:2045-2322
2045-2322
DOI:10.1038/s41598-023-35594-7