Regarding Bioanalysis Lasting a Few Minutes: Automated Cooling-SPME and Fast-GC for Urinary 2-Phenyl-2-Propanol Monitoring
An innovative SPME head space GC-MS method, in cooling mode, using a fully automated routine, was developed to detect 2-phenyl-2-propanol, a representative urinary metabolite of cumene. Following an acid hydrolysis and derivatization step with lowered quantities of reagents, acetic anhydride and pyr...
Gespeichert in:
Veröffentlicht in: | Toxics (Basel) 2024-10, Vol.12 (10), p.743 |
---|---|
Hauptverfasser: | , , , , , , , , , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | An innovative SPME head space GC-MS method, in cooling mode, using a fully automated routine, was developed to detect 2-phenyl-2-propanol, a representative urinary metabolite of cumene. Following an acid hydrolysis and derivatization step with lowered quantities of reagents, acetic anhydride and pyridine, a 30 μm polydimethylsiloxane SPME fiber was used to sample derivatized 2-phenyl-2-propanol, such as benzenemethanol,α,α-dimethyl-acetate, from the headspace. Performances of the method, optimized through experimental design, provide an LOD of 0.034 mg/L and an LOQ 0.10 mg/L, with a short sampling time necessary per sample. The method, developed on standard solutions, will be applied to both occupationally exposed and non-exposed populations. |
---|---|