Determination of ertapenem in plasma and ascitic fluid by UHPLC-MS/MS in cirrhotic patients with spontaneous bacterial peritonitis
Spontaneous bacterial peritonitis is a frequent severe complication in cirrhotic patients with ascites. Carbapenem antibiotics are currently the treatment of choice for patients with hospital-acquired or healthcare-related infections. However, there is limited evidence available on the efficacy of e...
Gespeichert in:
Veröffentlicht in: | Advances in laboratory medicine 2024-06, Vol.5 (2), p.173-180 |
---|---|
Hauptverfasser: | , , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Spontaneous bacterial peritonitis is a frequent severe complication in cirrhotic patients with ascites. Carbapenem antibiotics are currently the treatment of choice for patients with hospital-acquired or healthcare-related infections. However, there is limited evidence available on the efficacy of ertapenem in cirrhotic patients with spontaneous bacterial peritonitis. As a result, the pharmacokynetics and pharmacodynamics of this antibiotic are still unknown. The objective of this study was to develop and validate measurement procedures based on liquid chromatography-tandem mass spectrometry (UHPLC-MS/MS) to determine ertapenem concentrations in plasma and ascitic fluid.
Samples were pretreated by acetronile protein-precipitation. Chromatographic separation is performed on a C
reversed-phase Acquity
-UPLC
-BEH
column (2.1 × 100 mm id, 1.7 µm) using a non-linear gradient of water/acetonitrile containing 0.1 % of formic acid at a flow rate of 0.4 mL/min. Ertapenem and its internal standard (ertapenem-D
) are detected by tandem mass spectrometry using positive electrospray ionization and multiple reaction monitoring, and using 476.2 → 346.0/432.2 as mass transition for ertapenem and 480.2 → 350.0 for its internal standard.
No significant interferences or carry-over contamination were observed. Imprecisions, absolute relative bias, matrix effects and normalized recoveries were ≤14.5 %, ≤9.3 % (92.8-104.5) % and (98.8-105.8) %, respectively. Chromatographic measurement procedures were linear from (0.50-100) mg/L.
The measurement procedures based on UHPLC-MS/MS developed and validated in this study could be useful in pharmacokynetic and pharmacodynamic studies in subjects with liver cirrhosis who develop spontaneous bacterial peritonitis treated with ertapenem. |
---|---|