Innovative management of peritoneal catheter malfunction caused by omental wrapping: exploration the modified low-temperature plasma ablation blade in vivo and in vitro
Omental wrapping is a common complication of peritoneal dialysis, which manifests as catheter malfunction causing early termination of peritoneal dialysis. This study investigates the efficacy and safety of a Modified Low-Temperature Plasma Ablation (MLTPA) blade in treating omental-wrapped catheter...
Gespeichert in:
Veröffentlicht in: | Renal failure 2024-12, Vol.46 (2), p.2430375 |
---|---|
Hauptverfasser: | , , , , , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Omental wrapping is a common complication of peritoneal dialysis, which manifests as catheter malfunction causing early termination of peritoneal dialysis. This study investigates the efficacy and safety of a Modified Low-Temperature Plasma Ablation (MLTPA) blade in treating omental-wrapped catheter malfunction.
experiments using the MLTPA blade at the 5th power level were conducted to ablate omental tissues and catheters, evaluating ablation effects and potential catheter damage.
, nine beagles were modeled for catheter malfunction due to omental wrapping, randomized into MLTPA blade (
= 5) and the Gastroscopic Biopsy Forceps (GBF) groups (
= 4). Dialysate outflow velocity, dialysate residual volume, and complications were recorded.
experiments showed that the MLTPA blade effectively ablated omental tissue without damaging the catheter at the 5th power level.
, the MLTPA blade group achieved a 100% (5/5) catheter recanalization success rate, significantly increasing catheter outflow velocity (111.4 ± 5.1 ml vs. 3.0 ± 1.0 ml,
|
---|---|
ISSN: | 0886-022X 1525-6049 1525-6049 |
DOI: | 10.1080/0886022X.2024.2430375 |