Innovative management of peritoneal catheter malfunction caused by omental wrapping: exploration the modified low-temperature plasma ablation blade in vivo and in vitro

Omental wrapping is a common complication of peritoneal dialysis, which manifests as catheter malfunction causing early termination of peritoneal dialysis. This study investigates the efficacy and safety of a Modified Low-Temperature Plasma Ablation (MLTPA) blade in treating omental-wrapped catheter...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Renal failure 2024-12, Vol.46 (2), p.2430375
Hauptverfasser: Weng, Xiaoxue, Wang, Qiang, Qiu, Miaohua, Shen, Qingyi, Chen, Jiejian, Yan, Lingling, Yang, Jingbao, Huang, Ruojing, Wen, Wen, Yu, Guoqing
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:Omental wrapping is a common complication of peritoneal dialysis, which manifests as catheter malfunction causing early termination of peritoneal dialysis. This study investigates the efficacy and safety of a Modified Low-Temperature Plasma Ablation (MLTPA) blade in treating omental-wrapped catheter malfunction. experiments using the MLTPA blade at the 5th power level were conducted to ablate omental tissues and catheters, evaluating ablation effects and potential catheter damage. , nine beagles were modeled for catheter malfunction due to omental wrapping, randomized into MLTPA blade (  = 5) and the Gastroscopic Biopsy Forceps (GBF) groups (  = 4). Dialysate outflow velocity, dialysate residual volume, and complications were recorded. experiments showed that the MLTPA blade effectively ablated omental tissue without damaging the catheter at the 5th power level. , the MLTPA blade group achieved a 100% (5/5) catheter recanalization success rate, significantly increasing catheter outflow velocity (111.4 ± 5.1 ml vs. 3.0 ± 1.0 ml,  
ISSN:0886-022X
1525-6049
1525-6049
DOI:10.1080/0886022X.2024.2430375