Determination of Fat-Soluble Vitamins A, D3, and E in Infant Formula and Milk Powder Using High-Performance Liquid Chromatography with Photodiode Array Detection: Jordan Market as a Case Study
A new, simple, rapid, and sensitive reversed-phase High-Performance Liquid Chromatography-Photodiode Array (HPLC-PDA) method was developed and validated for the simultaneous analysis of fat-soluble vitamins A, D3, and E. The method required a simple sample preparation step of saponification with aqu...
Gespeichert in:
Veröffentlicht in: | Iranian journal of chemistry & chemical engineering 2020-07, Vol.39 (4), p.173-181 |
---|---|
Hauptverfasser: | , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | A new, simple, rapid, and sensitive reversed-phase High-Performance Liquid Chromatography-Photodiode Array (HPLC-PDA) method was developed and validated for the simultaneous analysis of fat-soluble vitamins A, D3, and E. The method required a simple sample preparation step of saponification with aqueous KOH and extraction with n-hexane. The method was validated in terms of linearity, accuracy, precision, stability, detection limits, and recovery. The method has the advantage of simultaneous determination of vitamins A, D3, and E in a short run time of 10 min. The method was applied successfully for the determination of vitamins A, D3, and E in some infant formula and milk powder from the Jordanian market. Based on this study, vitamin Content in all brands was within 90.0-410% of the labeled value. Vitamin D3 content in the studied brands was within 100.0-850%, while vitamin E content in all brands was less than 48%. The results showed large variation and discrepancy of most vitamins’ contents, which were not in good agreement with the manufacturer label value, though they were below toxic levels. |
---|---|
ISSN: | 1021-9986 1021-9986 |
DOI: | 10.30492/ijcce.2019.34728 |