Analytical performance of a single epitope B-type natriuretic peptide sandwich immunoassay on the Minicare platform for point-of-care diagnostics
Point-of-care B-type natriuretic peptide (BNP) testing with adequate analytical performance has the potential to improve patient flow and provide primary care givers with easy-to-use advanced diagnostic tools in the management of heart failure. We present the analytical evaluation of the Minicare BN...
Gespeichert in:
Veröffentlicht in: | Practical laboratory medicine 2019-05, Vol.15, p.e00119-e00119, Article e00119 |
---|---|
Hauptverfasser: | , , , , , , , , , , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Point-of-care B-type natriuretic peptide (BNP) testing with adequate analytical performance has the potential to improve patient flow and provide primary care givers with easy-to-use advanced diagnostic tools in the management of heart failure. We present the analytical evaluation of the Minicare BNP immunoassay under development on the Minicare I-20 platform for point-of-care testing. Analytical performance was evaluated using EDTA venous whole blood, EDTA plasma and capillary whole blood. Method comparison with a lab-testing system was performed using samples from 187 patients. Normal values were determined based on 160 healthy adults, aging from 19 to 70 years. Limit of blank (LoB), limit of detection (LoD) were determined to be 3.3 ng/L, 5.8 ng/L. Limit of quantitation (LoQ) in whole blood at 20% and 10% coefficient of variation (CV) was found |
---|---|