Leg occlusion of the Ovation Alto stent graft main body caused by twisting of the stent graft: A case report
The Ovation Alto stent graft is a unique endovascular abdominal aortic aneurysm repair device. A low-viscosity, radiopaque liquid polymer fills the rings located within the legs of the device. We reported a rare case of a patient diagnosed with an abdominal aortic aneurysm who was surgically treated...
Gespeichert in:
Veröffentlicht in: | Journal of vascular surgery cases and innovative techniques 2024-12, Vol.10 (6), p.101625, Article 101625 |
---|---|
Hauptverfasser: | , , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | The Ovation Alto stent graft is a unique endovascular abdominal aortic aneurysm repair device. A low-viscosity, radiopaque liquid polymer fills the rings located within the legs of the device. We reported a rare case of a patient diagnosed with an abdominal aortic aneurysm who was surgically treated using endovascular abdominal aortic aneurysm repair with an Ovation Alto stent graft. Polymer underfilling due to contralateral leg twisting of the stent graft’s main body resulted in limb occlusion. We highlighted the importance of avoiding the rotation of the Ovation Alto stent graft, thus preventing the potential twisting of this device within the sheath after the insertion of the stent graft’s main body. |
---|---|
ISSN: | 2468-4287 2468-4287 |
DOI: | 10.1016/j.jvscit.2024.101625 |