Determining the error in measuring the spherical concave mirror radius of curvature with a laser rangefinder
An instrument and technique for assessing errors of measuring optical surface radius of curvature with a laser rangefinder are presented. Errors of optical instrument alignment with a wave-front sensor are shown to influence the accuracy of measuring the mirror radius. Errors of the rangefinder-aide...
Gespeichert in:
Veröffentlicht in: | Kompʹûternaâ optika 2023-04, Vol.47 (2), p.246-250 |
---|---|
Hauptverfasser: | , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | An instrument and technique for assessing errors of measuring optical surface radius of curvature with a laser rangefinder are presented. Errors of optical instrument alignment with a wave-front sensor are shown to influence the accuracy of measuring the mirror radius. Errors of the rangefinder-aided technique for measuring the surface radius are estimated. A computer analysis shows that the developed scheme of misalignment measurement allows a relative error of 0.02 – 0.3 % to be attained for mirrors ranging in radius from 1 m to 10 m. The choice of the accuracy characteristics of the rangefinders used for measuring the optical surface radius of curvature is justified. |
---|---|
ISSN: | 0134-2452 2412-6179 |
DOI: | 10.18287/2412-6179-CO-1197 |