Transthoracic Vector Flow Imaging in Pediatric Patients with Valvular Stenosis – A Proof of Concept Study
Abstract Purpose Continuous wave Doppler ultrasound is routinely used to detect cardiac valve stenoses. Vector flow imaging (VFI) is an angle-independent real-time ultrasound method that can quantify flow complexity. We aimed to evaluate if quantification of flow complexity could reliably assess val...
Gespeichert in:
Veröffentlicht in: | Ultrasound international open 2021-08, Vol.7 (2), p.E48-E54 |
---|---|
Hauptverfasser: | , , , , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Abstract
Purpose
Continuous wave Doppler ultrasound is routinely used to detect
cardiac valve stenoses. Vector flow imaging (VFI) is an angle-independent
real-time ultrasound method that can quantify flow complexity. We aimed to
evaluate if quantification of flow complexity could reliably assess valvular
stenosis in pediatric patients.
Materials and Methods
Nine pediatric patients with echocardiographically
confirmed valvular stenosis were included in the study. VFI and Doppler
measurements were compared with transvalvular peak-to-peak pressure differences
derived from invasive endovascular catheterization.
Results
Vector concentration correlated with the catheter measurements
before intervention after exclusion of one outlier (r=−0.83,
p=0.01), whereas the Doppler method did not (r=0.49,
p=0.22). The change in vector concentration after intervention
correlated strongly with the change in the measured catheter pressure difference
(r=−0.86, p=0.003), while Doppler showed a tendency for
a moderate correlation (r=0.63, p=0.07).
Conclusion
Transthoracic flow complexity quantification calculated from
VFI data is feasible and may be useful for assessing valvular stenosis severity
in pediatric patients. |
---|---|
ISSN: | 2509-596X 2199-7152 |
DOI: | 10.1055/a-1652-1261 |