Monte Carlo dosimetry study of newly designed shielded applicators for intensity modulated brachytherapy of cervical and vaginal cancers
Introduction The utilization of metal shields in intensity-modulated brachytherapy (IMBT) enables the modulation of the dose, resulting in improved conformance to the tumor while simultaneously reducing the doses to organs at risk (OARs). Utilizing higher-energy sources like 60 Co in IMBT for cervic...
Gespeichert in:
Veröffentlicht in: | Discover. Oncology 2024-09, Vol.15 (1), p.494-10, Article 494 |
---|---|
Hauptverfasser: | , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Introduction
The utilization of metal shields in intensity-modulated brachytherapy (IMBT) enables the modulation of the dose, resulting in improved conformance to the tumor while simultaneously reducing the doses to organs at risk (OARs). Utilizing higher-energy sources like
60
Co in IMBT for cervical and vaginal cancers has consistently posed challenges. This study evaluates the dosimetric aspects of modified applicators designed for IMBT using
60
Co and
192
Ir sources.
Materials and Methods
GATE, a Geant4-based simulation code, was utilized to model and simulate four distinct applicators. The clinical applicators were redesigned to place the structure of the source tube and the shield while keeping the general characteristics unchanged. These shields were evaluated by calculating transmission factors (TFs) and the dose homogeneities were also determined.
Result
Transmission factors for the IMBT technique in redesigned intrauterine applicators and tungsten shields for iridium and cobalt sources were at least 12.8 and 65.4%, and these values were obtained for the intravaginal applicator at 0.2 and 7.0%, respectively. The dose homogeneities for all combinations of radionuclide-shield were within a 15% range of the non-IMBT applicators.
Conclusion
This study has quantitatively evaluated the dosimetric effect of tungsten shields in the IMBT technique for cervical and vaginal cancer using cobalt sources.
192
Ir compared to
60
Co resulted in higher effectiveness for the designed intrauterine and intravaginal shields. while implementing tungsten shields in the redesigned applicators against the
60
Co source may not offer complete protection, it does show promising results in reducing the dose to organs at risk. |
---|---|
ISSN: | 2730-6011 2730-6011 |
DOI: | 10.1007/s12672-024-01383-5 |