Canted seating of the Stryker Modular Dual Mobility liner within a Trident hemispherical acetabular shell

Abstract A 75-year-old woman who suffered a left femoral neck fracture underwent a left total hip arthroplasty using a Stryker Trident (Kalamazoo, MI) hemispherical acetabular shell and Modular Dual Mobility (MDM) metal liner. Post-operative radiographs demonstrated canted seating of the liner. The...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Arthroplasty today 2016-03, Vol.2 (1), p.19-22
Hauptverfasser: Eskildsen, Scott M., MD, Olsson, Erik C., MD, Del Gaizo, Daniel J., MD
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:Abstract A 75-year-old woman who suffered a left femoral neck fracture underwent a left total hip arthroplasty using a Stryker Trident (Kalamazoo, MI) hemispherical acetabular shell and Modular Dual Mobility (MDM) metal liner. Post-operative radiographs demonstrated canted seating of the liner. The patient was taken immediately back to the operating room where the acetabular liner appeared well seated superiorly but was in a canted position inferiorly. Removal and replacement was performed and post-operative radiographs demonstrated complete seating. Subsequent follow up at 6 months demonstrated good clinical function with no adverse radiographic findings. Canted seating is a potential complication of the MDM metal liner. Providers should be aware of potential incomplete seating inferiorly despite the superior portion of the liner being well seated.
ISSN:2352-3441
2352-3441
DOI:10.1016/j.artd.2015.05.003