First-in-human noninvasive left ventricular ultrasound pacing: A potential screening tool for cardiac resynchronization therapy
A screening tool to predict response to cardiac resynchronization therapy (CRT) could improve patient selection and outcomes. The purpose of this study was to investigate the feasibility and safety of noninvasive CRT via transcutaneous ultrasonic left ventricular (LV) pacing applied as a screening t...
Gespeichert in:
Veröffentlicht in: | Heart rhythm O2 2023-02, Vol.4 (2), p.79-87 |
---|---|
Hauptverfasser: | , , , , , , , , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | A screening tool to predict response to cardiac resynchronization therapy (CRT) could improve patient selection and outcomes.
The purpose of this study was to investigate the feasibility and safety of noninvasive CRT via transcutaneous ultrasonic left ventricular (LV) pacing applied as a screening test before CRT implants.
P-wave–triggered ultrasound stimuli were delivered during bolus dosing of an echocardiographic contrast agent to simulate CRT noninvasively. Ultrasound pacing was delivered at a variety of LV locations with a range of atrioventricular delays to achieve fusion with intrinsic ventricular activation. Three-dimensional cardiac activation maps were acquired via the Medtronic CardioInsight 252-electrode mapping vest during baseline, ultrasound pacing, and after CRT implantation. A separate control group received only the CRT implants.
Ultrasound pacing was achieved in 10 patients with a mean of 81.2 ± 50.8 ultrasound paced beats per patient and up to 20 consecutive beats of ultrasound pacing. QRS width at baseline (168.2 ± 17.8 ms) decreased significantly to 117.3 ± 21.5 ms (P |
---|---|