Difficult Ventilation in an Infant After Successful Intubation

Manufacturing defects in the connector of the endotracheal tube are not frequently encountered in emergency and planned intubations. In cases where they are encountered, however, they may cause partial or complete airway obstruction in sick infants with limited reserves, giving rise to a life-threat...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Turkish Journal of Anaesthesiology and Reanimation 2022-08, Vol.50 (4), p.303-305
Hauptverfasser: Evren Sahin, Kubra, Salman Onemli, Canan
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:Manufacturing defects in the connector of the endotracheal tube are not frequently encountered in emergency and planned intubations. In cases where they are encountered, however, they may cause partial or complete airway obstruction in sick infants with limited reserves, giving rise to a life-threatening situation following the intubation. For this reason, endotracheal tubes must be checked carefully before use. To this end, a stylet can be passed through the tube to check for a defect in the tube lumen or tube connector as part of a routine check of an endotracheal tube. This report features a patient who experienced a complete airway obstruction following intubation due to a manufacturing defect in the connector of the endotracheal tube.
ISSN:2667-677X
2667-6370
DOI:10.5152/TJAR.2021.21191