Calibration methods of new device for human blood pressure measurement
The paper presents two methods of determining calibration curve of a new device for blood pressure measurement. The device was developed at Wrocław University of Science and Technology. First method is based on parallel measurement of systolic and diastolic pressure measurement with use of reference...
Gespeichert in:
Veröffentlicht in: | EPJ Web of conferences 2019, Vol.201, p.6003 |
---|---|
Hauptverfasser: | , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | The paper presents two methods of determining calibration curve of a new device for blood pressure measurement. The device was developed at Wrocław University of Science and Technology. First method is based on parallel measurement of systolic and diastolic pressure measurement with use of reference device such as sphygmomanometer and researched new device with pneumatic sensor equipped with voltage type output. Obtained data (systolic p
s
and diastolic p
d
pressure, maximum u
s
and minimum u
d
voltage) was then used to determine individual pressure-voltage characteristic of the device, which can be represented as a linear equation. Second method is based on substitution of experimentally proved coefficient b with its analytical equivalent extracted from mathematical model of described pneumatic sensor. Described methods were verified experimentally and compared. Metrological parameters of the device were designated. |
---|---|