Traceable low-current measurements for a novel ionization gauge suitable as reference standard
A novel ionization gauge was developed where the electron beam is guided through the ionization space in a straight path. This design was chosen in order to develop a more stable ionization vacuum gauge suitable as reference standard in the range of 10−6 Pa to 10−2 Pa. The design requires a low elec...
Gespeichert in:
Veröffentlicht in: | Measurement. Sensors 2021-12, Vol.18, p.100202, Article 100202 |
---|---|
Hauptverfasser: | , , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | A novel ionization gauge was developed where the electron beam is guided through the ionization space in a straight path. This design was chosen in order to develop a more stable ionization vacuum gauge suitable as reference standard in the range of 10−6 Pa to 10−2 Pa. The design requires a low electron current resulting in even lower ion currents. For the purpose of a reference standard, these ultra-low currents in the range 1 pA–100 nA need to be measured in a traceable manner. For the first time, an ultrastable low-noise current amplifier was used for this purpose. |
---|---|