A Contactless Laser Doppler Strain Sensor for Fatigue Testing with Resonance-Testing Machine

In this article, a non-contact laser Doppler strain sensor designed for fatigue testing with the resonance-testing machine is presented. The compact sensor measures in-plane displacements simultaneously from two adjacent points using the principle of in-plane, laser-Doppler vibrometry. The strain is...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Sensors (Basel, Switzerland) Switzerland), 2021-01, Vol.21 (1), p.319
Hauptverfasser: Wang, Fangjian, Krause, Steffen, Hug, Joachim, Rembe, Christian
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:In this article, a non-contact laser Doppler strain sensor designed for fatigue testing with the resonance-testing machine is presented. The compact sensor measures in-plane displacements simultaneously from two adjacent points using the principle of in-plane, laser-Doppler vibrometry. The strain is computed from the relative displacements divided by the distance between these two points. The optical design, the mathematical model for estimating noise-limited resolution, the simulation results of this model, and the first measurement results are presented. The comparison of the measurement results of our sensor with the results of a conventional strain gauge shows that our design meets the measurement requirements. The maximum strain deviation compared to conventional strain gauges of the laser-Doppler extensometer is below 4×10-5 in all performed experiments.
ISSN:1424-8220
1424-8220