A polyethylene liner dissociation case report in depuy pinnacle cup. An impingement problem?

The modular components facilitate joint replacement surgery but are associated with potential complications not present in the previous components. We present a case of polyethylene dissociation due to rupture of the anti-rotational blocks in a total hip prosthesis two years after surgery. We think...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Archives of Clinical and Experimental Surgery 2019-06, Vol.8 (2), p.1-52
Hauptverfasser: Ayora, Alvaro, Gonzalez, Guillermo, Fernandez, Carlos
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:The modular components facilitate joint replacement surgery but are associated with potential complications not present in the previous components. We present a case of polyethylene dissociation due to rupture of the anti-rotational blocks in a total hip prosthesis two years after surgery. We think about the possible causes and risk factors. Despite its low incidence, the dissociation between polyethylene and the acetabular cup is a catastrophic complication. The femoroacetabular impingement seems to play a key role. [Arch Clin Exp Surg 2019; 8(2.000): 49-52]
ISSN:2146-8133
2146-8133
DOI:10.5455/aces.20190318055641