HTS library plate rejuvenation using a DMSO-rich atmosphere

Dry DMSO can rapidly pull water vapor out of the air due to its hygroscopic nature. This is a well-documented problem within drug discovery, particularly within high-throughput screening (HTS). This hydration is caused by atmospheric moisture being absorbed each time a compound library is used. This...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:SLAS technology 2024-12, Vol.29 (6), p.100204, Article 100204
Hauptverfasser: Hughes, Fraser, Cookson, Alex, Tamaki, Fabio, Bailey, Christopher, Gray, David W., Wrobel, Karolina, Cookson, Kirsty, Bell, Steve, Tarver, Gary
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:Dry DMSO can rapidly pull water vapor out of the air due to its hygroscopic nature. This is a well-documented problem within drug discovery, particularly within high-throughput screening (HTS). This hydration is caused by atmospheric moisture being absorbed each time a compound library is used. This effect becomes increasingly pronounced when a compound library is used routinely. The result of this hydration is a change to both the total volume of solution and the concentration of sample still in solution. This can result in a large amount of variability in the measured biological activity of a sample depending on the library usage. In this paper, we show the detrimental effects the hydration of sample libraries has on the reproducibility of biological data and present a novel way to remove it from HTS library plates. Our approach involves creating a DMSO-rich environment, created by placing anhydrous DMSO in compound storage pods purged with nitrogen, and incubating library plates in this environment for up to 3 days. Quantification via evaporative light scattering detection (ELSD) showed that removing water greatly increased the molarity of solutions, with a greater effect being seen for compounds with poor solubility. We also demonstrated how this approach can restore the inhibitory activity of stock solutions of compounds (pIC50) of samples containing ∼30 % water from >30 µM to sub-micromolar after moisture removal. This method improves the reliability of tested compounds in HTS by potentially saving pharmaceutical companies hundreds of thousands of dollars in screening campaigns and increasing the quality of data. [Display omitted]
ISSN:2472-6303
2472-6311
2472-6311
DOI:10.1016/j.slast.2024.100204