Relationship between the ablation index, lesion formation, and incidence of steam pops
Background The ablation index (AI) is reported to be useful for a durable pulmonary vein isolation (PVI). However, there have been no studies investigating the relationship between the power, contact force (CF), AI, and steam pops. Methods Using an in vitro model, ablation energy was delivered until...
Gespeichert in:
Veröffentlicht in: | Journal of arrhythmia 2019-08, Vol.35 (4), p.636-644 |
---|---|
Hauptverfasser: | , , , , , , , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Background
The ablation index (AI) is reported to be useful for a durable pulmonary vein isolation (PVI). However, there have been no studies investigating the relationship between the power, contact force (CF), AI, and steam pops.
Methods
Using an in vitro model, ablation energy was delivered until a steam pop occurred and the time to the steam pop and AI when the steam pop occurred were measured. The experiment was performed with a combination of various powers (20, 30, 40, and 50 W) and contact forces (CFs) (10, 30, and 50 g) 20 times for each setting. The analysis consisted of two protocols. The first protocol was a comparison between the ablation power and several parameters under the same CF (10, 30, and 50 g). The second protocol was a comparison between the CF and several parameters under the same power (20, 30, 40, and 50 W). The correlation between the lesion formation and ablation parameters was evaluated.
Results
The AI value when steam pops occurred varied depending on the ablation settings. All AI median values were |
---|---|