Evaluation of deformation and fracture rates for nickel-titanium rotary instruments according to the frequency of clinical use

To evaluate the deformation and fracture rates for ProTaper Universal (PTU) nickel-titanium rotary instruments according to the frequency of clinical use. A total of 619 PTU instruments (S1, S2, F1, F2, and F3) that have been used in the clinic by a single endodontist were collected over a period of...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:European Oral Research 2018-05, Vol.52 (2), p.89-93
Hauptverfasser: Yilmaz, Ayca, Gokyay, Sitki Selcuk, Daglaroglu, Rustu, Karagoz Kucukay, Isil
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:To evaluate the deformation and fracture rates for ProTaper Universal (PTU) nickel-titanium rotary instruments according to the frequency of clinical use. A total of 619 PTU instruments (S1, S2, F1, F2, and F3) that have been used in the clinic by a single endodontist were collected over a period of 4 years. These instruments were grouped on the basis of one to three (Group A), four to six (Group B) and seven to nine (Group C) clinical uses (one canal = one use). All instruments were evaluated by a blinded investigator under a stereomicroscope at 15×-45× magnification for the presence of deformation and fracture. The overall rates of deformation and fracture were 10% and 1.2%, respectively. The deformation and fracture rates for the S2, F1, and F2 instruments showed no significant differences among groups. However, fracture rate for S1 instruments in Group A was significantly higher than for those in Group B (p=0.025) and Group C (p=0.004). In Group C, the S1 instruments showed a significantly higher deformation rate compared with the S2 (p=0.04), F1 (p=0.008) and F2 (p=0.049) instruments; there were no other significant differences within groups. Under the conditions of the current study, frequency of use seemed to influence the deformation rates of PTU rotary instruments. Except S1, these instruments could be used without any fracture or deformation in up to 9 clinical cases by an experienced endodontist.
ISSN:2630-6158
2149-2352
2651-2823
2149-4592