Correction of Thermal Deviations of Fabry-Perot Resonator Based Measurements of Specific Gases in Millimeter Wave Bands

Due to the thermal expansivity of the material used in the Fabry-Perot resonator mirrors, the resonator cavity length can change and this might therefore have an impact on the resonant frequency during high-resolution spectroscopy measurements. Based on measurements and simulations, this paper discu...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Radioengineering 2012-04, Vol.21 (1), p.459-463
Hauptverfasser: J. Libich, P. Dvorak, P. Piksa, S. Zvanovec
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:Due to the thermal expansivity of the material used in the Fabry-Perot resonator mirrors, the resonator cavity length can change and this might therefore have an impact on the resonant frequency during high-resolution spectroscopy measurements. Based on measurements and simulations, this paper discusses the influences of temperature on the precise determination of gas attenuation measured in a Fabry-Perot resonator. Several measures to mitigate such influence and to correct the measured results were tested. A correction method for the measured data was proposed.
ISSN:1210-2512