Speaking valve with integrated biomimetic overpressure release and acoustic warning signal

Speaking valves enable tracheostomy patients to speak naturally. However, improper use may cause dangerous overpressure, leading to severe complications or even patient’s death. We address this life-threatening issue by creating a biomimetic speaking valve, which incorporates an integrated overpress...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Scientific reports 2024-11, Vol.14 (1), p.26655-11, Article 26655
Hauptverfasser: Knorr, N., Auth, P., Kruppert, S., Stahl, C. A., Lücking, K. M., Tauber, F., Speck, T.
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:Speaking valves enable tracheostomy patients to speak naturally. However, improper use may cause dangerous overpressure, leading to severe complications or even patient’s death. We address this life-threatening issue by creating a biomimetic speaking valve, which incorporates an integrated overpressure valve that automatically opens when reaching critical pressure levels. To enhance safety, we integrated a whistle module to provide an audible alert for medical staff. Fundamental research on the Utricularia vulgaris trapdoor inspired our abstracted valve form. Through a comprehensive analysis using generalized linear mixed-effect models, we examined various membrane parameter effects on the function of the biomimetic overpressure valve. This enabled us to adjust the valve’s opening pressure to cater to patient’s unique requirements, thus potentially saving lives by applying a solution from nature.
ISSN:2045-2322
2045-2322
DOI:10.1038/s41598-024-77595-0