Theoretical and Experimental Study of Heterodyne Phase-Sensitive Dispersion Spectroscopy with an Injection-Current-Modulated Quantum Cascade Laser
We report the theoretical and experimental study of calibration-free heterodyne phase-sensitive dispersion spectroscopy (HPSDS) in the mid-infrared using a direct current modulated mid-infrared quantum cascade laser (QCL). The modulation of QCL current at several hundred MHz or higher generates the...
Gespeichert in:
Veröffentlicht in: | Sensors (Basel, Switzerland) Switzerland), 2020-10, Vol.20 (21), p.6176 |
---|---|
Hauptverfasser: | , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | We report the theoretical and experimental study of calibration-free heterodyne phase-sensitive dispersion spectroscopy (HPSDS) in the mid-infrared using a direct current modulated mid-infrared quantum cascade laser (QCL). The modulation of QCL current at several hundred MHz or higher generates the synchronous frequency and intensity modulation of the QCL emission. An analytical model of the phase of the beat note signal in HPSDS is derived by considering the absorption and dispersion processes and incorporating the QCL modulation parameters. In the experiment, a 4.5 μm QCL modulated at 350 MHz was used to measure N2O at 200 Torr in a 10 cm gas cell. The N2O concentrations inferred from the analytical model were compared with the nominal values to show good agreement over the concentration range of 189−805 ppm with a standard deviation |
---|---|