Measurement Uncertainty Evaluation of Digital Modulation Quality Parameters: Magnitude Error and Phase Error

In digital modulation quality parameters traceability, the Error Vector Magnitude, Magnitude Error and Phase Error must be traced, and the measurement uncertainty of the above parameters needs to be assessed. Although the calibration specification JJF1128-2004 Calibration Specification for Vector Si...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Hauptverfasser: Zhan, Zhiqiang, Xia, Junwen, Wang, Yunling, Yu, Lei
Format: Tagungsbericht
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:In digital modulation quality parameters traceability, the Error Vector Magnitude, Magnitude Error and Phase Error must be traced, and the measurement uncertainty of the above parameters needs to be assessed. Although the calibration specification JJF1128-2004 Calibration Specification for Vector Signal Analyzers is published domestically, the measurement uncertainty evaluation is unreasonable, the parameters selected is incorrect, and not all error terms are selected in measurement uncertainty evaluation. This article lists formula about magnitude error and phase error, than presents the measurement uncertainty evaluation processes for magnitude error and phase errors.
ISSN:2261-236X
2274-7214
2261-236X
DOI:10.1051/matecconf/20166104002