A Multiple Emergency Ventilator as Backup Solution in Pandemic: A Specifically Designed and Dimensioned Device

Goal: To provide a Multiple Emergency Ventilator (MEV) as backup in case of shortage of ICU ventilators and for use in camp hospitals. Methods: MEV provides the same oxygen mixture and peak inspiratory pressure (PIP) to 10 patients. These specifications were fixed: i) gas supply and plugs to double-...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:IEEE open journal of engineering in medicine and biology 2022, Vol.3, p.41-46
Hauptverfasser: Baselli, Giuseppe, Fiore, Gianfranco, Casella, Francesco, Cinquemani, Simone, Vigano, Roberto, Pesenti, Antonio, Zanella, Alberto
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:Goal: To provide a Multiple Emergency Ventilator (MEV) as backup in case of shortage of ICU ventilators and for use in camp hospitals. Methods: MEV provides the same oxygen mixture and peak inspiratory pressure (PIP) to 10 patients. These specifications were fixed: i) gas supply and plugs to double-limb intubation sets compatible to existing systems; ii) fluid-dynamics with no pressure drop and almost complete patients' uncoupling; iii) individual monitoring of inspiratory and expiratory pressures and flows and control of their timing; iv) easy stocking, transport, installation with self-supporting pipes. Results: A Bell-Jar System (BJS) design permitted to safely fix PIP based on Archimedes' law. The main distribution line was based on 2" stainless steel pipes assuring the required mechanical properties and over-dimensioned for fluidics. The Windkessel of the BJS and pipeline dead-volumes is 75.65 L and in the worst case of the instantaneous demand of 5 L by 10 patients (0.5 L each) shows an adiabatic PIP drop limited to -6.18%, confirming the needed uncoupling. Consequently, patients' asynchrony is permitted as needed by pressure-controlled volume-guaranteed and assisted-ventilation. Conclusions: Although MEV is proposed as a backup system, its features may cover the whole set of ventilation modes required by ICU ventilation.
ISSN:2644-1276
2644-1276
DOI:10.1109/OJEMB.2022.3152673