Abnormally high Patient State Index associated with epicardial pacing: a case report

Background Anesthesiologists monitor electroencephalography (EEG) intraoperatively to maintain adequate depth of anesthesia. However, the EEG signal is affected by noise and interference. The SedLine® is a brain function monitor with which the Patient State Index (PSI) is calculated. In this study,...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:JA clinical reports 2022-05, Vol.8 (1), p.35-35, Article 35
Hauptverfasser: Yonezawa, Hiroki, Maeda, Takuma, Takise, Yoshiaki, Morinaga, Masahiro, Ohnishi, Yoshihiko
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:Background Anesthesiologists monitor electroencephalography (EEG) intraoperatively to maintain adequate depth of anesthesia. However, the EEG signal is affected by noise and interference. The SedLine® is a brain function monitor with which the Patient State Index (PSI) is calculated. In this study, we report abnormally high PSI values associated with epicardial pacing during open heart surgery. Case presentation A 50-year-old man was scheduled for total arch replacement. Atrial demand pacing was started before weaning from cardiopulmonary bypass. The PSI increased from 30 to 80 soon after the start of pacing, and the EEG waveform showed spikes synchronized with the pacing. As the pacing output was lowered, the spikes on the EEG attenuated and disappeared, and the PSI decreased to < 40. When the pacing output was increased again, the spikes recurred, and the PSI increased again. Conclusions Pacemaker spikes may cause contamination of the EEG, resulting in abnormally high PSI values.
ISSN:2363-9024
2363-9024
DOI:10.1186/s40981-022-00521-8