Triboelectric touch sensor for position mapping during total hip arthroplasty
In this research, a triboelectric nanogenerator (TENG) was utilized to determine if a pressure-based sensor could detect bearing friction in a total hip arthroplasty (THA) and detect the contact of specific areas during ROM checks. The pressure-based sensor shows capability to sense bearing friction...
Gespeichert in:
Veröffentlicht in: | BMC research notes 2020-08, Vol.13 (1), p.395-395, Article 395 |
---|---|
Hauptverfasser: | , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | In this research, a triboelectric nanogenerator (TENG) was utilized to determine if a pressure-based sensor could detect bearing friction in a total hip arthroplasty (THA) and detect the contact of specific areas during ROM checks.
The pressure-based sensor shows capability to sense bearing friction. In more detail, the TENG embedded in four different sides of the trial exhibits up to 1 V from peak-to-peak. Moreover, these flexible touch sensors with TENG describes a peak signal in output voltage which should lead to extremely sensitive detection of bearing friction induced by the THA. |
---|---|