A new digitised screen test for strabismus measurement

Our study presents a digitised tangent screen test for ocular motility analysis according to the Harms and Hess tests (measurement of the squint angle in all fields of vision). This test uses an image beamer to display the tangent screen, a position sensor to measure the patient’s head orientation,...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Zeitschrift für medizinische Physik 2023-05, Vol.33 (2), p.182-191
Hauptverfasser: Oltrup, Theo, Bende, Marvin, Henseling, Celine, Bende, Thomas, Leitritz, Martin A, Ulrich Bartz-Schmidt, Karl
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:Our study presents a digitised tangent screen test for ocular motility analysis according to the Harms and Hess tests (measurement of the squint angle in all fields of vision). This test uses an image beamer to display the tangent screen, a position sensor to measure the patient’s head orientation, and a distance sensor to measure the fixation distance. Digital measurement of head orientation allows for a test procedure that eliminates the conventional requirement for a light pointer in the patient’s hand. Thus, the digital screen test is presented, and the uncertainty of the measurement system is evaluated. A mathematical relationship was given between the measured squint angles, as well as the angle of diagnostic gaze direction, and the influence quantities on their measurement uncertainty. The individual uncertainties resulted from deviations in the measured values by the position and distance sensors, the calibration of the projection image of the beamer in length units, and the finite image resolution of the beamer. The individual standard uncertainties of the influence quantities were determined. The combined standard measurement uncertainties of the squint and gaze direction angles were given based on the model equation of the error propagation law at the tangent table according to Harms at a test distance of 2.5 m. The patient’s uncertainty contribution to the mobility analysis was not considered. The combined standard uncertainty of the measurement system (coverage factor k = 2 for 95% confidence level) for the squint angle is ≤ 0.43° for the angle of diagnostic gaze direction ≤ 3.13° at the test distance of 2.5 m. The individual standard uncertainties of the influence quantities on the angles are (k = 1): 1.55°/1.01° (horizontal/vertical angle of the position sensor), 0.19° (distance sensor), 0.06° (calibration of the projection image of the beamer), and 0.02° (image resolution of the beamer). The maximum valid test distance of the digital screen test is 3.8 m. The digital screen test is compact and can be used at different locations. Compared to the traditional test, the time required for examination via the digitised test is less; additionally, its documentation is simplified. The measurement uncertainty of the diagnostic gaze direction angle is dominated by the sensor drift of the position sensor in the horizontal direction (yaw angle) and is due to the sensor technology. However, this drift error does not affect the squint angle measurement r
ISSN:0939-3889
1876-4436
DOI:10.1016/j.zemedi.2022.07.001