Radiation Dose Reduction in Different Digital Radiography Systems: Impact on Assessment of Defined Bony Structures in Bearded Dragons ( Pogona vitticeps )

Three different digital detector systems were used to study the effect of a defined radiation dose reduction on the image quality of digital radiographs in bearded dragons ( ). A series of radiographs of seven bearded-dragon cadavers with a body mass ranging from 132 g to 499 g were taken in dorsove...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Animals (Basel) 2023-05, Vol.13 (10), p.1613
Hauptverfasser: Steiner, Natalie, Ludewig, Eberhard, Tebrün, Wiebke, Pees, Michael
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:Three different digital detector systems were used to study the effect of a defined radiation dose reduction on the image quality of digital radiographs in bearded dragons ( ). A series of radiographs of seven bearded-dragon cadavers with a body mass ranging from 132 g to 499 g were taken in dorsoventral projection. The digital systems employed included two computed radiography systems (CR) (one system with a needle-based and one with a powdered-based scintillator) and one direct radiography system (DR). Three levels of the detector dose were selected: A standard dose (defined based on the recommended exposure value of the CR , D/100%), a half dose (D/50%), and a quarter dose (D/25%). Four image criteria and one overall assessment were defined for each of four anatomic skeletal regions (femur, rib, vertebra, and phalanx) and evaluated blinded by four veterinarians using a predefined scoring system. The results were assessed for differences between reviewers (interobserver variability), radiography systems, and dosage settings (intersystem variability). The comparison of the ratings was based on visual grading characteristic (VGC) analysis. Dose reduction led to significantly lower scores in all criteria by every reviewer, indicating a linear impairment of image quality in different skeletal structures in bearded dragons. Scores did not differ significantly between the different systems used, indicating no advantage in using a computed or direct radiography system to evaluate skeletal structures in bearded dragons. The correlation was significant ( ≤ 0.05) for interobserver variability in 100% of the cases, with correlation coefficients between 0.50 and 0.59. While demonstrating the efficacy of the use of digital radiography in bearded dragons and the similar quality in using different computed or direct radiography systems, this study also highlights the importance of the appropriate level of detector dose and demonstrates the limits of post-processing algorithm to compensate for insufficient radiation doses in bearded dragons.
ISSN:2076-2615
2076-2615
DOI:10.3390/ani13101613