Procedimiento para el cálculo de las especificaciones de exactitud en el rango de temperatura de 700 °c a 1000 °c de un horno tubular
We developed a process for calibrating a Terrigeno Industries furnace tube, model ESP series 462 in a 700 ° C to 1000 ° C range, in order to determine the accuracy and the confidence interval of the temperature in the equipment when heat treatments are made on materials studied. Using the standard m...
Gespeichert in:
Veröffentlicht in: | Scientia et technica 2011, Vol.2 (48), p.169-172 |
---|---|
Hauptverfasser: | , , , |
Format: | Artikel |
Sprache: | spa |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | We developed a process for calibrating a Terrigeno Industries furnace tube,
model ESP series 462 in a 700 ° C to 1000 ° C range, in order to determine the
accuracy and the confidence interval of the temperature in the equipment
when heat treatments are made on materials studied.
Using the standard metrological analysis for the calculation of uncertainties
in direct measurements, we conclude that the contribution to the expanded
uncertainty, by the uncertainty type B due to accuracy specifications of the
furnace, is 2% of the maximum temperature programmed in the 700 ºC at 1000
°C range.
Se desarrolló un proceso para la calibración de un horno tubular marca Industrias Terrígeno, Modelo ESP serie 462 en un rango de 700 °C a 1000 °C, con el fin de determinar la exactitud y el intervalo de confianza de la temperatura programada en el equipo cuando se efectúan tratamientos térmicos a materiales estudiados.
Utilizando el análisis metrológico estándar para el cálculo de incertidumbres en medidas directas, se concluye que la contribución a la incertidumbre expandida, por la incertidumbre tipo B debido a especificaciones de exactitud del horno es del 2% de la temperatura máxima programada en el rango de 700 ºC a 1000 ºC. |
---|---|
ISSN: | 0122-1701 |