Mercury Films on Commercial Carbon Screen-Printed Devices for the Analysis of Heavy Metal Ions: a Critical Evaluation
The suitability of mercury films on commercial screen‐printed electrodes for the analysis of heavy metal ions is critically tested for the particular case of Pb(II)‐ions. Although determination is possible by anodic stripping voltammetry with a reasonable detection limit (8.9 µg L−1), important draw...
Gespeichert in:
Veröffentlicht in: | Electroanalysis (New York, N.Y.) N.Y.), 2015-06, Vol.27 (6), p.1345-1349 |
---|---|
Hauptverfasser: | , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | The suitability of mercury films on commercial screen‐printed electrodes for the analysis of heavy metal ions is critically tested for the particular case of Pb(II)‐ions. Although determination is possible by anodic stripping voltammetry with a reasonable detection limit (8.9 µg L−1), important drawbacks are noticed as a consequence of the heterogeneous deposition of mercury on the rough surface of screen‐printed devices. |
---|---|