Performance characterisation of a real-time fiber dosimetry system using radiophotoluminescent glasses
We report the first demonstration of a real-time fiber-optic-coupled dosimetry system using Ag-activated phosphate glasses based on radiophotoluminescent (RPL) phenomena. The performance characterisation using real-time fiber RPL glass dosimetry is compared with that of real-time fiber plastic scint...
Gespeichert in:
Veröffentlicht in: | Japanese Journal of Applied Physics 2018-10, Vol.57 (10), p.106402 |
---|---|
1. Verfasser: | |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | We report the first demonstration of a real-time fiber-optic-coupled dosimetry system using Ag-activated phosphate glasses based on radiophotoluminescent (RPL) phenomena. The performance characterisation using real-time fiber RPL glass dosimetry is compared with that of real-time fiber plastic scintillation dosimetry. The inevitable build-up phenomena occurring in Ag-activated RPL glasses both during and after X-ray exposure were measured. The real-time RPL curves for the X-ray exposure region and the build-up region are analytically evaluated and fitted to the second-order polynomial functions and the sum of three exponentials, respectively. In addition, some resolved problems of real-time fiber RPL glass dosimetry are discussed. |
---|---|
ISSN: | 0021-4922 1347-4065 |