High-speed ultrahigh-resolution spectral domain optical coherence tomography using high-power supercontinuum at 0.8 µm wavelength
We demonstrated high-speed ultrahigh-resolution (UHR) optical coherence tomography (OCT) in the 800 nm wavelength region. A high-power coherent supercontinuum (SC) and a high-speed line scan camera were used to construct a spectral domain OCT. The axial resolution was 3.1 µm in air and 2.3 µm in tis...
Gespeichert in:
Veröffentlicht in: | Applied physics express 2015-07, Vol.8 (8), p.82501 |
---|---|
Hauptverfasser: | , , , , |
Format: | Artikel |
Sprache: | eng |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | We demonstrated high-speed ultrahigh-resolution (UHR) optical coherence tomography (OCT) in the 800 nm wavelength region. A high-power coherent supercontinuum (SC) and a high-speed line scan camera were used to construct a spectral domain OCT. The axial resolution was 3.1 µm in air and 2.3 µm in tissue. The dependence of sensitivity on the SC power and A-scan rate was examined. For the A-scan rate of 70 kHz, the sensitivity of 104 dB was achieved for the SC power higher than 60 mW. High-speed in vivo UHR-OCT imaging was demonstrated for zebrafish embryo and swimming medaka. |
---|---|
ISSN: | 1882-0778 1882-0786 |
DOI: | 10.7567/APEX.8.082501 |