New Method of Estimating Effective Energy for X-ray CT Scanners

Because the exposure dose in X-ray computed tomography examinations is sometimes difficult to determine, it is important to be able to estimate the dose for these examinations. The effective energy of the X-ray CT scanner is required to estimate exposure dose. Although the half-value-layer(HVL) meth...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Japanese Journal of Radiological Technology 2005/03/20, Vol.61(3), pp.385-391
Hauptverfasser: NAGASHIMA, HIROYUKI, SUNAGA, SHINICHI, MITOME, TOYOHISA, KOYA, EIICHI, TAKEI, HIROYUKI, MATSUBARA, KUNIO, OGURA, TOSHIHIRO, SHIRAISHI, AKIHISA, SHIMOSEGAWA, MASAYUKI, HOSHINA, MASAO, IGARASHI, HITOSHI
Format: Artikel
Sprache:eng ; jpn
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:Because the exposure dose in X-ray computed tomography examinations is sometimes difficult to determine, it is important to be able to estimate the dose for these examinations. The effective energy of the X-ray CT scanner is required to estimate exposure dose. Although the half-value-layer(HVL) method has been used to calculate effective energy, it is not an easy method. This paper proposes a technique by which effective energy can be easily calculated. Certain details were found to cause change in effective energy, and the ratio (inner-metal center-air ratio : IMCAR) between air dose and dose in fixing the metallic pipe in the isocenter of an X-ray CT scanner was necessary. The IMCAR from a different X-ray CT scanner was required, and, when effective energy was calculated, it showed an error of less than 0.7% for the half-value-layer method. The effect of this error on dose estimation was slight (0.4%). This technique is useful, because effective energy can easily be calculated with a high degree of accuracy.
ISSN:0369-4305
1881-4883
DOI:10.6009/jjrt.KJ00003326699