Partial Electrical Reset of CT Irradiation on Implantable Cardiac Devices: Relationship between Reset and Tube Voltage, Tube Current, and Rotation Time
The aim of this study was to investigate the relationship between partial electrical reset (PER) and CT scan parameters (tube voltage, tube current, rotation time, and product of tube current and rotation time in mAs). A cardiac resynchronization therapy pacemaker (Insync 8040, Medtronic Inc., Tokyo...
Gespeichert in:
Veröffentlicht in: | Japanese Journal of Radiological Technology 2011/06/25, Vol.67(6), pp.648-653 |
---|---|
Hauptverfasser: | , , , , , , |
Format: | Artikel |
Sprache: | eng ; jpn |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | The aim of this study was to investigate the relationship between partial electrical reset (PER) and CT scan parameters (tube voltage, tube current, rotation time, and product of tube current and rotation time in mAs). A cardiac resynchronization therapy pacemaker (Insync 8040, Medtronic Inc., Tokyo) and 320 area detector CT scanner (Aquilion ONE, Toshiba medical systems, Otawara, Japan) with volume scan were used. The pacemaker was put in DDD mode. The PERs were interpreted using both the programmer's wave forms and error messages. The exposure was repeated 5 times per CT setting. The pacemaker was placed on the anterior wall and upper side of a chest phantom. Each CT scan was performed using the following parameters: tube voltage of 80, 100, 120, and 135 kV; tube current of 50-550 mA; and rotation time of 0.35-1.5 s. PERs were observed at 100, 120, and 135 kV, and more PERs were observed as the tube voltage increased. The PER tube current decreased as the rotation time was increased. In contrast, the PER tube current and rotation time product (mAs) increased as the rotation time was increased. More specifically, the radiation dose rate was the affected factor of the PERs. To avoid PER of pacemakers, CT scan parameters with lower radiation dose rates (low rather than high tube current and rotational time) is recommended. In conclusion, our results will help with CT scans of patients who have implantable cardiac devices (included pacemakers and cardioverter defibrillators). |
---|---|
ISSN: | 0369-4305 1881-4883 |
DOI: | 10.6009/jjrt.67.648 |