Development of a novel gauge filled with resin for X-ray CT
X-ray computed tomography (CT) systems have been widely used for three-dimensional (3D) measurements in the engineering field. Therefore, verifying X-ray CT systems with full traceability is necessary to ensure accurate measurements. Measurement errors of X-ray CT systems are evaluated using a refer...
Gespeichert in:
Veröffentlicht in: | E-journal of Nondestructive Testing 2022-03, Vol.27 (3) |
---|---|
Hauptverfasser: | , , , , |
Format: | Artikel |
Sprache: | eng |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | X-ray computed tomography (CT) systems have been widely used for three-dimensional (3D) measurements in the engineering field. Therefore, verifying X-ray CT systems with full traceability is necessary to ensure accurate measurements. Measurement errors of X-ray CT systems are evaluated using a reference gauge whose geometry is calibrated using 3D measuring systems with higher accuracy. Thus, this study developed and characterized a novel reference gauge for X-ray CT systems, comprising spheres embedded in a cylindrical resin. |
---|---|