Energy Response of an Optically Stimulated Luminescence Dosemeter to Low Energy Continuous X-rays
To investigate the energy response of an optically stimulated luminescence dosemeter (OSLD) with low energy photons, OSLDs were irradiated by low energy continuous X-rays which have various energy distributions. The effective energy of the X-ray was in the range 10.0-14.5keV, and homogeneity factor...
Gespeichert in:
Veröffentlicht in: | Hoken butsuri 2006, Vol.41(4), pp.249-255 |
---|---|
Hauptverfasser: | , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | To investigate the energy response of an optically stimulated luminescence dosemeter (OSLD) with low energy photons, OSLDs were irradiated by low energy continuous X-rays which have various energy distributions. The effective energy of the X-ray was in the range 10.0-14.5keV, and homogeneity factor was in the range 0.5-0.9. As a result, at 11.0keV, the OSLD-to-ionization chamber ratio was maintained at almost a constant level from 1.03 at a 0.5 to 1.08 at a 0.7 homogeneity factor. However, the OSLD tended to overestimate measured value with an increase of effective energy. Probably, this error can be eliminated by calculating an appropriate correction factor. |
---|---|