Validated HPTLC Method for Simultaneous Estimation of Atenolol and Aspirin in Bulk Drug and Formulation
This paper describes a new, simple, precise, and accurate HPTLC method for simultaneous estimation of Atenolol and Aspirin as the bulk drug and in tablet dosage forms. Chromatographic separation of the drugs was performed on aluminum plates precoated with silica gel 60 F254 as the stationary phase a...
Gespeichert in:
Veröffentlicht in: | ISRN analytical chemistry 2012-02, Vol.2012 (2012), p.1-5 |
---|---|
Hauptverfasser: | , |
Format: | Artikel |
Sprache: | eng |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | This paper describes a new, simple, precise, and accurate HPTLC method for simultaneous estimation of Atenolol and Aspirin as the bulk drug and in tablet dosage forms. Chromatographic separation of the drugs was performed on aluminum plates precoated with silica gel 60 F254 as the stationary phase and the solvent system consisted of n-butanol : water : acetic acid (8 : 2 : 0.2 v/v/v). Densitometric evaluation of the separated zones was performed at 235 nm. The two drugs were satisfactorily resolved with RF values 0.23±0.02 and 0.72±0.02 for Atenolol and Aspirin, respectively. The accuracy and reliability of the method was assessed by evaluation of linearity (100–600 ng/spot for Atenolol and Aspirin), precision (intraday % RSD was 0.48–1.03 and interday % RSD was 0.68–1.14 for Atenolol, and intraday % RSD was 0.61–1.03 and interday % RSD was 0.69–1.04 for Aspirin), accuracy (100.17±0.32 for Atenolol and 99.73±0.51 for Aspirin), and specificity in accordance with ICH guidelines. |
---|---|
ISSN: | 2090-7311 2090-732X 2090-732X |
DOI: | 10.5402/2012/609706 |