In Vivo Intraocular Pressure Monitoring during Microincision Vitrectomy with and without Active Control of Infusion Pressure
Purpose To evaluate intraocular pressure (IOP) fluctuation during vitrectomy, we directly monitored IOP in vivo using 2 vitrectomy machines with or without constant infusion pressure monitoring and control. Methods Among 61 eyes of 61 consecutive patients, 32 were assigned to the Accurus system (gro...
Gespeichert in:
Veröffentlicht in: | European journal of ophthalmology 2017-09, Vol.27 (5), p.601-606 |
---|---|
Hauptverfasser: | , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Purpose
To evaluate intraocular pressure (IOP) fluctuation during vitrectomy, we directly monitored IOP in vivo using 2 vitrectomy machines with or without constant infusion pressure monitoring and control.
Methods
Among 61 eyes of 61 consecutive patients, 32 were assigned to the Accurus system (group 1) and 29 were assigned to the Constellation system (group 2) in this prospective case series. The IOP fluctuations were evaluated during routine vitrectomy procedures.
Results
The initial IOP before vitrectomy was 20.3 ± 2.4 mm Hg in group 1 using a conventional vented gas forced infusion system and 20.0 ± 0.0 mm Hg in group 2 using active IOP control at 20 mm Hg (p = 0.532). However, the average IOP change during core vitrectomy was -8.6 ± 4.3 mm Hg in group 1 and -0.8 ± 1.1 in group 2 (p |
---|---|
ISSN: | 1120-6721 1724-6016 |
DOI: | 10.5301/ejo.5000956 |