Chlorine activation by N 2 O 5 : simultaneous, in situ detection of ClNO 2 and N 2 O 5 by chemical ionization mass spectrometry
We report a new method for the simultaneous in situ detection of nitryl chloride (ClNO2) and dinitrogen pentoxide (N2O5) using chemical ionization mass spectrometry (CIMS). The technique relies on the formation and detection of iodide ion-molecule clusters, I(ClNO2)− and I(N2O5)−. The novel N2O5 det...
Gespeichert in:
Veröffentlicht in: | Atmospheric measurement techniques 2009-05, Vol.2 (1), p.193-204 |
---|---|
Hauptverfasser: | , , |
Format: | Artikel |
Sprache: | eng ; jpn |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | We report a new method for the simultaneous in situ detection of nitryl chloride (ClNO2) and dinitrogen pentoxide (N2O5) using chemical ionization mass spectrometry (CIMS). The technique relies on the formation and detection of iodide ion-molecule clusters, I(ClNO2)− and I(N2O5)−. The novel N2O5 detection scheme is direct. It does not suffer from high and variable chemical interferences, which are associated with the typical method of nitrate anion detection. We address the role of water vapor, CDC electric field strength, and instrument zero determinations, which influence the overall sensitivity and detection limit of this method. For both species, the method demonstrates high sensitivity (>1 Hz/pptv), precision (~10% for 100 pptv in 1 s), and accuracy (~20%), the latter ultimately determined by the nitrogen dioxide (NO2) cylinder calibration standard and characterization of inlet effects. For the typically low background signals ( |
---|---|
ISSN: | 1867-8548 1867-8548 |
DOI: | 10.5194/amt-2-193-2009 |