An UPLC-MS detection method for the quantification of five antibiotics in human plasma
An UPLC-MS detection method for the quantification of amikacin, flucloxacillin, meropenem, penicillin G and vancomycin was developed and validated. The calibration curves were found to be linear from 0.47 to 50 mg/l for amikacin, 1.28 to 135 mg/l for flucloxacillin, 0.75 to 80 mg/l for meropenem, 0....
Gespeichert in:
Veröffentlicht in: | Bioanalysis 2015-09, Vol.7 (18), p.2321-2329 |
---|---|
Hauptverfasser: | , , , , , , , , |
Format: | Artikel |
Sprache: | eng |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | An UPLC-MS detection method for the quantification of amikacin, flucloxacillin, meropenem, penicillin G and vancomycin was developed and validated.
The calibration curves were found to be linear from 0.47 to 50 mg/l for amikacin, 1.28 to 135 mg/l for flucloxacillin, 0.75 to 80 mg/l for meropenem, 0.38 to 80 mg/l for penicillin G and 0.73 to 80 mg/l for vancomycin. Between- and within-run accuracy was ranged between 85 and 115%. Between and within imprecision expressed as CV was within 15%.
The validated method was successfully applied to a PK study in very preterm and small for gestational age infants treated for nosocomial sepsis and/or meningitis. |
---|---|
ISSN: | 1757-6180 1757-6199 |
DOI: | 10.4155/bio.15.121 |