A device for dried blood microsampling in quantitative bioanalysis: overcoming the issues associated blood hematocrit

A cross-laboratory experiment has been performed on a novel dried blood sampler in order to investigate whether it overcomes issues associated with blood volume and hematocrit (HCT) that are observed when taking a subpunch from dried blood spot samples. An average blood volume of 10.6 μl was absorbe...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Bioanalysis 2015-04, Vol.7 (6), p.653-659
Hauptverfasser: Spooner, Neil, Denniff, Philip, Michielsen, Luc, De Vries, Ronald, Ji, Qin C, Arnold, Mark E, Woods, Karen, Woolf, Eric J, Xu, Yang, Boutet, Valérie, Zane, Patricia, Kushon, Stuart, Rudge, James B
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:A cross-laboratory experiment has been performed on a novel dried blood sampler in order to investigate whether it overcomes issues associated with blood volume and hematocrit (HCT) that are observed when taking a subpunch from dried blood spot samples. An average blood volume of 10.6 μl was absorbed by the samplers across the different HCTs investigated (20-65%). No notable change of volume absorbed was noted across the HCT range. Furthermore, the variation in blood sample volumes across six different laboratories was within acceptable limits. The novel volumetric absorptive microsampling device has the potential to deliver the advantages of dried blood spot sampling while overcoming some of the issues associated with the technology.
ISSN:1757-6180
1757-6199
DOI:10.4155/bio.14.310